Emetteurs thermocouple en-tête

Smart Temperature Transmitters

C$180.78

Article# TX13

Remises de volume disponibles
Calendrier de remise sur volume
Quantité Prix unitaire
1-4 C$180.78
5-9 C$171.74
10-24 C$168.13
25-49 C$160.89
50-99 C$157.28
100+ C$153.66

En stock

Délai de livraison (si pas en stock): 6 semaines
Smart Temperature Transmitters TX13 TX13
  • Entrées thermocouple K, J, N, E, T, R, S et mV
  • Configuration via touches
  • 56 plages préprogrammées prêts-à-utiliser
  • Fonction Trim - annuler erreur de capteur
  • Usure programmable avec indicateur de panne du capteur
  • Isolation galvanique

TX13 Options de modèles

Voir tous les modèles

Modifiez ces spécifications pour commander un modèle différente. Toutes les combinaisons ne sont pas valides. Les options compatibles avec les sélections précédentes seront en gras.

*Les options en surbrillance ne sont pas compatibles. Veuillez sélectionner une combinaison différente.

Input Type
Input Class
Input Ranges
Output Type
Output Class
Supply Voltage Type
Programability


Description

Le TX13 est un émetteur "en-tête" intelligent qui accepte des capteurs de température thermocouple et convertit la sortie du capteur sur une plage configurée à un signal de transmission standard industriel  (4 à 20 mA). Une touche simple permet l'utilisateur de sélectionner le type de TC, la direction d'usure, des plages fixes et des points de Trim de 4 et 20 mA. L'émetteur incorpore la dernière technologie numérique pour assurer une performance précise sans dérive. Si nécessaire, la plage désirée peut être précisée lors de la commande, ce qui élimine le besoin de configuration par l'utilisateur. Si la plage n'est pas précisée alors l'émetteur sera envoyé avec la plage de défaut de 0 à 1000°C Type K.

CARACTERISTIQUES
Isolation: contrôlé à 250 Vcc
Usure du capteur: sortie convertisseur descendant ou ascendant
Soudure froide:
    Plage: -40 à 85°C 
    Précision: ±0,5°C
    Dépistage: ±0,05°C/°C
Stabilité:
    Décalée: 0,1°C/°C
    Portée: 0,05°C/°C
Sortie
Type de sortie: 2 fils, 4 à 20 mA loupe de courant
Plage de sortie: 4,0 à 20,0 mA
Connexion sortie: borne à vis
Sortie maximum: 21,5 mA (en condition usure haute)
Sortie minimum: 3,8 mA (en condition usure basse)
Précision: (sortie mA /2000) ou 5 uA (le plus haut s'applique)
Effet tension de boucle: 0,2 uA/V
Drift de température: 1 uA/°C
Charge maximum: [(alimentation V-10)/20]K? (Exemple: 700? @ 24V)
Temps de mise à jour: 500 mS
Temps de réponse: 1 s
Temps de démarrage: 4 s (1 sortie < 4 mA pendant démarrage)
Temps de réchauffement: 1 minute pour atteindre précision complète
Alimentation: 10 à 30 Vcc
Environnement
Plage de température ambiante: -40 à 85°C
Stockage de température ambiante: -50 à 90°C
Plage d'humidité ambiante: 10 à 90% HR  sans condensation
Dimensions physiques: Dia 43 mm; H 21 mm 
Poids: 31 g (encapsulé)
Standard
EMC - BS EN 61326 :1998: équipement électrique pour le contrôle de mesures et utilisation en laboratoire
ANNEXE A: tests d'immunité nécessaires pour équipement conçu pour utilisation dans des locaux industriels
loading all models
loading all reviews
Rated 5 out of 5 by from Great buy Excellent device for tight applications. I put it inside a 3" junction box and fit perfectly.
Date published: 2016-02-15
Rated 3 out of 5 by from Beginner I like the transmitter that you guys have, as it's very versatile. I am new to the field and I'm teaching myself setup and calibration. I thought that it would have been nice to know that I would have to expect to do that in advance and with more direction. The device certainly needed trimmed to the (-100C / 100C) range of the T type thermocouple. I thought that it would be a lot closer to the 4mA & 20mA output than it was. I had to buy a calibrator and learned how to operate the new machine and was able to trim that device to exactly (-100C / 100C) to 4mA & 20mA, and that was fun, and the device was easy to configure. It was so far off though it surprised me, it seems the true zero and span for the T type was actually set to something other than (-100C / 100C) so it's not plug and play, and maybe they never are. I'm a beginner, and now maybe a novice! What I've learned I will be able to apply to many things in our facility and I certainly like the easy configuration that your product offers, now that I know how it all works. The Best School is the School of Hard Knocks!
Date published: 2016-03-18
  • y_2024, m_4, d_17, h_5
  • bvseo_bulk, prod_bvrr, vn_bulk_3.0.41
  • cp_1, bvpage1
  • co_hasreviews, tv_0, tr_2
  • loc_en_CA, sid_TX13, prod, sort_[SortEntry(order=HELPFULNESS, direction=DESCENDING), SortEntry(order=POSITIVE_FEEDBACK, direction=DESCENDING), SortEntry(order=SUBMISSION_TIME, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 3ms
  • REVIEWS, PRODUCT
loading all QA

What is the country of origin for your TX13 transmitters?

Asked by: judib63
Thank you for the inquiry. TX13 origin is Great Britain. Best regards OMEGA
Answered by: NorwalkAE_RS
Date published: 2024-03-27

Could you please share reliability data (FIT/MTBF) for the TX13-TX-SCALED? Thank you

Asked by: RM2021
Hi RM2021, Thank you for the inquiry. Unfortunately we do not offer requested data on the requested product. Best regards OMEGA
Answered by: NorwalkAE_RS
Date published: 2021-10-27

To use with Omega SA2C-K, type K TC with -50 to 200C range: -50 to to 200 is not one of the preconfigured ranges for TX13. What is the PN to order if we want the transmitter come with -50 to 200 reconfiguration?

Asked by: emali53
Thank you for the inquiry. You can add factory configuration option to a new TX13 with part# TX13-TX-SCALED. When placing order add scaling detail to description on PO copy. Else, if placing order via web site, add scaling details to Special Instructions box in line item details in shopping cart. Best regards OMEGA
Answered by: NorwalkAE_RS
Date published: 2023-08-13

Can you use any type of TC on this TX? ex. J or T if it only states K on the device?

Asked by: NCJOe
Thank you for your inquiry NCJOe. TX13 is compatible with K, J, N, E, T, R, S Thermocouples.
Answered by: Application Engineer SP
Date published: 2022-08-10
  • y_2024, m_4, d_17, h_5CST
  • bvseo_bulk, prod_bvqa, vn_bulk_3.0.41
  • cp_1, bvpage1
  • co_hasquestionsanswers, tq_4
  • loc_en_CA, sid_TX13, prod, sort_[SortEntry(order=NUM_ANSWERS, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 7ms