Skip to search Skip to main content Skip to navigation

Appelez-nous au 1-888-826-6342

Nous avons remarqué que vous veniez de . Visitez votre site local pour connaître les offres régionales et l'assistance en direct.

Contrôleurs de débit massique et volumétrique pour 14 gaz

Mass Flow Controllers with 20+ Gas Select Function

C$4,268.20

Article# FMA-2610A

En stock

Délai de livraison (si pas en stock): 3 semaines
Mass Flow Controllers with 20+ Gas Select Function FMA-2610A FMA-2610A
  • Choix de 14 Gaz: He, O2 , Neon, N2 O, N2, Air, Argon, CO, CO2 , Méthane, Ethane, Propane, Butane, H2
  • Affichage simultané de la pression, température, des débits volumétrique et massique (FMA-2600)
  • Commandes directes par bouton-poussoirs
  • Etalonnage NIST en option
  • Gammes pleine échelle de 0,5CCM à 1000LPM+
  • Temps de réponse typique de 100ms
  • Turndown Ratio of 100:1 Typical
  • Fonctionne dans toutes les positions
  • Précision de ±1% pleine échelle
  • Interface RS232 en standard

FMA-2610A Options de modèles

Voir tous les modèles

Modifiez ces spécifications pour commander un modèle différente. Toutes les combinaisons ne sont pas valides. Les options compatibles avec les sélections précédentes seront en gras.

*Les options en surbrillance ne sont pas compatibles. Veuillez sélectionner une combinaison différente.

Media Type
Flow Range
Accuracy
Output Signal
Wetted Materials
Max Operating Pressure
Max Process Temperature
Process Connection Type:
Process Connection Size:
Features
Category
Communications Protocol
Display


Description

Les contrôleurs de débit en flux laminaire FMA-2600 et FVL-2600 comprennent une vanne proportionnelle couplée à un débitmètre. Les mesures sont effectuées dans la zone de flux laminaire et le régulateur PID intégré positionne la vanne selon les points de consigne choisis.

Ces appareils robustes, fonctionnant en boucle fermée peuvent recevoir une commande de valeur de consigne 0-5V. Ils peuvent aussi recevoir un signal de commande par l'interface RS232. Indépendamment des tensions de consigne, les contrôleurs peuvent être configurés pour sortir un ou deux signaux tels que débit massique, température, débit ou pression. Ces signaux proviennent des capteurs de pression différentielle et absolue et de température qui servent à déterminer le débit. Il en résulte que cet appareil d'un emploi facile et à réponse rapide est caractérisé par une excellente linéarité et la capacité d"afficher plusieurs paramètres de procédé. Toutes les données de procédé, ainsi que la commande à distance du point de consigne sont simultanément disponibles par l'interface RS232. On peut gérer jusqu'à 26 contrôleurs par la même interface série. Les contrôleurs de gaz sont proposés de 0,5cm³/mn à 1m³/mn pleine échelle. (Les contrôleurs volumétriques sont plus économiques car ils ne nécessitent pas de capteur de pression absolue. Ils sont fréquemment utilisés dans les applications où la pression est faible)

Les prix des modèles les plus courants sont donnés dans le tableau "Pour commander". Utilisez la table de "Construction d'une référence" ci-dessous pour déterminer le prix des autres modèles.

Pour obtenir les caractéristiques complètes de ce produit, consultez la notice commerciale par le lien en bas de cette page.
loading all models
loading all reviews
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 4ms
  • REVIEWS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;
loading all QA

Is there an option to prevent electrical noise via user-selectable mitigation level?

Asked by: JMADV
Thank you for your inquiry. Please compose an email to flow@omega.com and desribe the issue or concern about electrical noise the flow controller may be exposed or currently being exposed and have our Design Team respond.
Answered by: Application Engineer JC
Date published: 2023-10-09

I am interested in using Modbus protocol via RS232 to read data from this instrument. What register address correspond to what type of data (ie Mass flow, Volumetric flow, Pressure and Temperature)?

Asked by: AI70
Thank you for your inquiry. Please refer to the manual that is available for MODBUS communication protocol and is found on the mass flow meter Product Page, FMA-1600A.
Answered by: Applicaton Engineer JJC
Date published: 2024-02-05

Is this tolerant of 100ppm SO2 or NO2?

Asked by: AGCQ
Thank you for your inquiry. This controller can handle the corrosive gases in question, yet, you will need to purchase the '-LSS' version that is a custom part number which will be chemically compatible.
Answered by: Applicaton Engineer JJC
Date published: 2023-12-01

Is there another version on serial communication protocol for the FMA-2600 series that does not involve the HyperTerminal?

Asked by: shep25
Thank you for your inquiry! Unfortunately, we do not offer any software with these units and with HyperTerminal not being included in newer Windows versions another program will be needed. Putty may be used as an alternate if HyperTerminal is not available
Answered by: Application Engineer JM
Date published: 2022-06-13

We have a FMA-2616A Mass Flow controller. Can we change the upstream valve configuration to downtream ourselves? Is there an introduction to the relevant operations? Is there a greater risk of damage to the instrument in such operations?

Asked by: herc
Thank you for your inquiry. The mass flow controller can work in vacuum conditions and it is recommended to have it returned via the Customer Care Group at CSReturns@omega.com and request to have the existing controller be used in vacuum conditions. The valve must be physically placed downstream of the flow to control accurately.
Answered by: Application Engineer JC
Date published: 2023-04-17

Do you have the distance from the center of the gas inlet hole to the edge of the device? I can attach a diagram showing which dimensions I need if there is a way to do that

Asked by: AGCQ
Thank you for your inquiry about the mass flow meter. Please forward that picture or drawing to flow@omega.com for the distance you described.
Answered by: Application Engineer JC
Date published: 2023-01-27

Are the FMA-2600 series mas flow controllers rated for vacuum use?

Asked by: joe simon
Thank you for your inquiry about the mass flow controllers. Yes, it can, yet, the valve must be ordered with the valve on the down stream side of the controller and will have a suffix, '-DS.' to reflect. If not found on the website, please email flow@omega.com to request a quote for a controller with the '-DS.'
Answered by: Application Engineer JC
Date published: 2023-04-17
  • y_2024, m_3, d_27, h_9CST
  • bvseo_bulk, prod_bvqa, vn_bulk_3.0.38
  • cp_1, bvpage1
  • co_hasquestionsanswers, tq_7
  • loc_en_CA, sid_FMA-2610A, prod, sort_[SortEntry(order=NUM_ANSWERS, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 44ms
  • QUESTIONS, PRODUCT
Nylon Body, TUBE ID 1/16" TO MALE 10-32UNF
C$11.09
0 In Stock
Nylon Body, TUBE ID 1/8" TO MALE 10-32UNF
C$11.09
0 In Stock
Polypropylene Body, TUBE ID 1/16" TO MALE 10-32UNF
C$11.09
4 In Stock
USB TO MALE 8 PIN MINI DIN CABLE 6
C$215.82
1 In Stock
Accessories
Replacement 1.8M (6') 8 pin male Mini-DIN single ended, cable connector
C$83.78
0 In Stock
1.8M (6') 8 pin male Mini-DIN double ended, cable connector
C$101.89
0 In Stock
7.62M (25') 8 pin male Mini-DIN single ended, cable connector
C$160.58
0 In Stock
7.62M (25') 8 pin male Mini-DIN double ended, cable connector
C$190.47
0 In Stock
1.8M (6') 8 pin male Mini-DIN to DB9 female adapter
C$96.47
3 In Stock
1.8 M (6') 8-pin male right-angle Mini_DIN single ended connector cable
C$101.89
0 In Stock
24 Vdc Universal power supply
C$150.85
0 In Stock
Accessories for FLR-1600/FLV-4600 Series
Multi-drop box
C$470.34
0 In Stock
fr-CA
fr
Enregistrer la configuration