hot point™ Dry Block Probe Calibrator, Single and Muli-Well, for Thermocouples and RTD's, Models CL900-(*) and CL950-(*)

hot point™ Dry Block Probe Calibrator

C$7,612.27

Article# CL900-220

En stock

Délai de livraison (si pas en stock): 6 semaines
hot point™ Dry Block Probe Calibrator CL900-220 CL900-220
  • Calibrates Thermocouples and RTD’s in Seconds
  • Highly Accurate 0.9°C (1.5°F) & Exceptional Stability (0.3°F) for Precision Calibrations
  • Completely Portable for Factory, Laboratory and Instrument Shops
  • Interchangeable Test Wells and CL950 Built-In 5 Port Well Design Provides Customer Flexibility

CL900-220 Options de modèles

Voir tous les modèles

Modifiez ces spécifications pour commander un modèle différente. Toutes les combinaisons ne sont pas valides. Les options compatibles avec les sélections précédentes seront en gras.

*Les options en surbrillance ne sont pas compatibles. Veuillez sélectionner une combinaison différente.

Supply Voltage
Process Temperature Unit
Process Temperature Min
Process Temperature Max
Process Temperature Range
Accuracy Value
Stability
Style
Communications Protocol
Additional Features


OMEGA’s new hot point® calibrators allow calibration of thermocouple and RTD probes quickly and easily with accuracy to 0.9°C (1.5°F). An integral RTD sensor assures high measurement accuracy and repeatability. A built-in microprocessor-based PID digital controller assures exceptional stability. Setpoint and actual temperature are displayed simultaneously. Temperature can be set in one-degree increments from 22°C (40°F) above ambient to 482°C (900°F).

The CL900 calibrator accepts a variety of standard and metric optional inserts which are available in 4" and 6" depths and for probe diameters from 1/8" to 3/8". An undrilled insert is also available. The CL900 comes complete with a 1/4" diameter, 6" deep insert well and insert block tongs. Additional inserts can be ordered separately as required for other diameter probes. 6" depth inserts are used for probes 7" and longer; 4" depth inserts are used for probes 5" and longer.

The CL950 calibrator features a built-in well with 5 ports sized to accommodate a 1/16", 1/8", 3/16", and two 1/4" diameter probes. The multi-port design allows for the use of a calibrated or NIST-traceable probe and meter to be used as a standard while calibrating other probes. The 5 port design also allows for calibration of more than one probe at a time, and eliminates the need for separate insert wells.

Inserts for CL900 Series
Model No. Description
CL901 Insert, 1/8 dia. x 4" deep test well
CL901-M Metric insert, 2 dia. x 101.6 mm deep test well
CL902 Insert, 1/8 dia. x 6" deep test well
CL902-M Metric insert, 2 dia. x 152.4 mm deep test well
CL903 Insert, 3/16 dia. x 4" deep test well
CL903-M Metric insert, 3 dia. x 101.6 mm deep test well
CL904 Insert, 3/16 dia. x 6" deep test well
CL904-M Metric insert, 3 dia. x 152.4 mm deep test well
CL905 Insert, 1/4 dia. x 4" deep test well
CL905-M Metric insert, 4.5 dia. x 101.6 mm deep test well
CL906 Insert, 1/4 dia. x 6" deep test well
CL906-M Metric insert, 4.5 dia. x 152.4 mm deep test well
CL907 Insert, 5/16 dia. x 4" deep test well
CL907-M Metric insert, 6 dia. x 101.6 mm deep test well
CL908 Insert, 5/16 dia. x 6" deep test well
CL908-M Metric insert, 6 dia. x 152.4 mm deep test well
CL909 Insert, 3/8 dia. x 4" deep test well
CL910 Insert, 3/8 dia. x 6" deep test well
CL911 Insert, undrilled
Specifications
(All specifications are based on the test probes being in contact with the bottom of the test well)

Temperature Range: Ambient +22 to 482°C (+40 to 900°F)
Operating Ambient Temperature Range: 5°C to 38°C (40°F to 100°F)
Accuracy: (1°F resolution)
CL900 with 6" test well inserts: ±1.5°F ±1 LSD.
CL900 with 4" test well inserts: ±3°F ±1 LSD.
CL950: ±1.5°F, ±1 LSD.
Control Stability: ±0.3°F or better
Temperature Uniformity (within the bottom 1" of the test well):
CL900:
with 6" test well inserts: 0.3%
with 4" test well inserts: ±0.4%.
CL950: ±0.3%
Max. Probe Immersion:
CL900:
with 6" test well inserts: 150 mm (6")
with 4" test well inserts: 100 mm (4").
CL950: 150 mm (6")
Control: Microprocessor-based PID digital controller with data retention
Power: 110-125 Vac, 50/60 Hz, 1050 watts; 220 Vac, 50/60 Hz, 1050 watts
Dimensions: 300 H x 216 W x 311 mm D (11.75 x 8.5 x 12.25")
Weight: 10.1 kg (22.3 lb)

CL900-110 hot point™ dry block calibrator
CL900-110 hot point™ dry block calibrator, with DSS-460-K benchtop thermometer, calibrating a single thermocouple probe, shown smaller than actual size.
multi-well accepts up to 5 probes at one time
CL950 multi-well accepts up to 5 probes at one time and comes standard with 1/16, 1/8, 3/16 and two 1/4" diameter holes.
multi-well accepts up to 5 probes at one time
CL900 accepts any insert listed in the table above.
loading all models
loading all reviews
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getReviews, 3ms
  • REVIEWS, PRODUCT
  • bvseo-msg: The resource to the URL or file is currently unavailable.;
loading all QA

I have type J thermocouples with brazed junction, can I calibrate with a dry well calibrator

Asked by: Anonymous
Thank you for your inquiry. As long as the thermocouple is the proper diameter and meet the minimum depth/length requirement, it can be used in this well.
Answered by: Application Engineer CB
Date published: 2022-12-10
  • y_2024, m_3, d_28, h_10CST
  • bvseo_bulk, prod_bvqa, vn_bulk_3.0.38
  • cp_1, bvpage1
  • co_hasquestionsanswers, tq_1
  • loc_en_CA, sid_CL900-220, prod, sort_[SortEntry(order=NUM_ANSWERS, direction=DESCENDING)]
  • clientName_omega
  • bvseo_sdk, java_sdk, bvseo-4.0.0
  • CLOUD, getContent, 16ms
  • QUESTIONS, PRODUCT
fr-CA
fr
Enregistrer la configuration