- Câble isolé minéral tolérance Classe 1
- Pression & isolation des sondes testées individuellement
- Gaine Acier inoxydable, Inconel® ou SuperOMEGACLAD™XL
- Diamètres de 1/16 to 1/4 "
- Fil toronné PFA 1 mètre (40"): 20 AWG pour 1/4" et 3/16" diam. ext., 24 AWG pour 1/8" et 1/16" diam. ext. sondes
- Des fils plus longs sont disponibles, veuillez consulter le tableau constructeur de référence ci-dessous
TJ36-CAIN-18U-6 Options de modèles
Voir tous les modèlesModifiez ces spécifications pour commander un modèle différente. Toutes les combinaisons ne sont pas valides. Les options compatibles avec les sélections précédentes seront en gras.
* Les options en surbrillance ne sont pas compatibles. Veuillez sélectionner une combinaison différente ou cliquez sur le bouton CONFIGURER pour créer la vôtre.
Thermocouple Type
Process Temperature Max
Process Temperature Range
Sensor Application
Sheath Material
Junction Type
Probe Diameter
Probe Length
Cable Insulation
Wire Gauge
Cable Length
Termination Connection Type
Color Code Standard
Additional Feature
Category
Process Temperature Unit
Process Temperature Min
Description
Les sondes à jonction renforcée OMEGA™ offrent une solution de raccordement pratique aux fils recouverts de PFA. La longueur de la jonction et du ressort de réduction de contrainte est de 66 mm (2.6"). Le diameter de la jonction de transition est de 1/4" pour les sondes de 1/16 et 1/8" de diamètre et de 3/8" pour les sondes de 3/16 et 1/4" de diamètre.Cliquez ici si vous désirez de l'assistance en choisissant une jonction thermocouple.
Rated 5 out of
5
by
Jim S from
Meets our process requirements
Product always ships in a timely manner and projected ship dates are met
Date published: 2023-01-16
Rated 4 out of
5
by
Chris_H from
Good usability but probe has surface scratches.
Good construction overall, and good response measuring water. My only critique is that the probe seems to have very apparent machining/grinding marks on it, perhaps from its manufacturing process. Don't expect it to be glass smooth.
Date published: 2022-10-21
Rated 5 out of
5
by
Nick I from
Well Made TC
This is a very well made thermocouple. The sheath is machined to high precision, and the response to temperature change is fairly quick.
Date published: 2022-03-25
Rated 5 out of
5
by
MichaelG from
Perfect
High quality. Fits our needs. Direct replacement for our equipment.
Date published: 2022-03-07
Rated 5 out of
5
by
Ryan Falconer Racing from
Excellent product
Thermocouples work great and are as advetised. We have been purchasing from Omega on and off for many decades as our need for thermocouples is sporadic. But when we need them, we go to Omega.
Date published: 2022-01-13
Rated 4 out of
5
by
Jeff Talk from
Customer service lacking
Called customer service and was promptly directed to the website first and then if I have any issues to call back. Product is awesome though.
Date published: 2022-01-03
Rated 5 out of
5
by
ESU1 from
Great Temp Sensors!
Great, high quality, long lasting K-Type Temp Sensors!
Date published: 2022-01-02
Rated 5 out of
5
by
Anonymous from
Great probe for temperature chamber
Very easy to set up, the rigidity of the probe allows in to stay it place close to the sample under test but it malleable enough to be bendable by hand. Highly recommend it.
Date published: 2021-03-11
Rated 3 out of
5
by
Rich64 from
Functional yes but improvement possibility
Ordered a SS k-type first but had some oxidation on probe. Followed up with a CAXL k-type and still have oxidation on probe. Both were used in the same SS protection tube. Ambient is a O2 rich environment using gas heated from 500 to 1700F. Not sure if exotic T/Cs would be a better choice.
Date published: 2020-12-22
Rated 5 out of
5
by
Mwgtech from
Product worked exactly what I needed it for.
Product was exactly what I needed. Product was of the high quality i expected.
Date published: 2020-12-08
Rated 5 out of
5
by
Laurel M from
Awesome replacement for mercury thermometers.
This probe is amazing! The fact that you can customize the materials and size to your specific needs is wonderful! We will never use mercury thermometers for our distillations again!
Date published: 2020-08-08
Rated 4 out of
5
by
C M R from
Lab Technician
This thermocouple is light and accurate in temperature.
Date published: 2020-05-28
Rated 4 out of
5
by
C M R from
Technician
Thermocouple is very accurate and light so it could be hold by a small clamp. Just make sure that the gauge of the wire is the right size for your project.
Date published: 2020-05-26
can these probes be bent? will bending them effect the readings?
Hello and thank you for contacting Omega. Yes, we can confirm that these probes are bendable to 90 degrees and that the bend will not affect the readings. We suggest a bending radius of 0.33” or larger. Please contact our Tech Support if you have any further questions.
Date published: 2021-11-08
How to determine accuracy of TJ36-? Not sure if listed special limits apply. Multiple configs: K or J, 316SS, varied sheath diameter & length, standard wire/length, grounded, SMPW-M, single & dual probes
Thank you for your inquiry SpecialStandard. The accuracy (tolerance) is determine by the Special Limits of Error according to the ASTM E230-ANSI MC 96.1. Please check the following link in page H-2 for the thermocouple tolerances for each thermocouple: https://assets.omega.com/landing-pages/colorcodes/tc_colorcodes.pdf. The tolerance only applies to probes.
Date published: 2024-02-11
Am I able to install these into an exhaust stream with temperatures up to 1450 F without a thermowell?
Hi Nick. That should be no problem although the if you want the added protection a thermowell with a threaded bushing would work.
Date published: 2020-08-18
I am looking at a T-type immersion, 304 SS, 1/8" dia, grounded TC. But looks like it is giving me only 40" cable length option. On picking up a 120" or 240" length it gives error saying compatibility issues. Please advise
Thank you for your inquiry. Please use the 'Configure Now' for custom lengths.
Date published: 2021-03-03
While the measurement point measures 1000 degrees, can there be a heat transfer to the transient point? In transient point, are there any protection for this?
Hi Gull,
Thank you for this inquiry. There is no protection built in for the transition point. Alternate is to provide enough length for the sheath so that heat transfer across the tubing is small enough to be safe.
Best regards
OMEGA
Date published: 2021-01-23
Do the ungrounded TCs have any electrical isolation rating?
Thank you for the inquiry. We do not offer an electrical isolation rating for our ungrounded thermocouples.
Best regards,
OMEGA
Date published: 2023-09-10