The HX200 is a high accuracy calibrated dew point or humidity transmitter. The remote sensor is designed to work at temperatures as high as 200°C (392°F). This high accuracy sensor is also built to withstand up to 750 psi. These transmitters are perfect for applications such as compressed air, dryers and environmental chambers.
Specifications Dew Point: -60 to 40°C (-76 to 104°F) Relative Humidity: 0 to 95% RH (non-condensing) unit displays dew point or humidity, as well as temperature Maximum Sensor Pressure Rating: 750 psi (51.7 bar) Sensor Body Material: 304 SS Maximum Operating Temperature: 200°C (392°F) (at 0% RH) Mesh Filter Top: Stainless Steel Single Output: 4 to 20 mA loop powered (relative humidity or dew point) Dew Point Accuracy: ±1.0°C from -20 to 40°C (-4 to 104°F); ±3.0°C below -20°C (-4°F) Relative Humidity Accuracy: ±1% RH from 15 to 40°C (59 to 104°F), 0 to 90% RH (non-condensing); ±2% RH below 15°C (59°F) or above 40°C (104°F) Display: 2 line LCD Resolution: 0.1 Power: 4 to 20 mA loop powered, 2-wire 24 Vdc (typical) 12 to 28 Vdc Dimensions (Enclosure): 88.9 L x 63.5 W mm (3.5 x 2.5") Mounting Center: 78.7 mm (3.1") Dimensions of Probe: Length: 10.16 cm (4") OD: 1.27 cm (0.5") ½ NPT Fitting Weight: 1.8 kg (4 lb)
Rated 5 out of
5 by
Dean Friesen from
Excellent product.Easy to incorporate in our system, operates flawlessly.
Date published: 2017-12-05
Rated 5 out of
5 by
humiditymeasurement from
can be connected to daqgot this for measuring inlet temperature and RH. can be connected to daq
Date published: 2022-09-12
Rated 5 out of
5 by
Copper from
works great in extreme environmentsThis instrument is more accurate than all of the other brands we have tried. It works great even through a high vibration environment where others fail
Date published: 2020-11-12
Rated 5 out of
5 by
rs1022 from
Works Great!Just finished installing yesterday, easy set up. Being used in Dew Point mode, and switched over via internal dip switched. This unit is loop powered, so it will require a 24vdc separate power source.
Date published: 2018-08-21
Rated 4 out of
5 by
Fibreman12345 from
Good productUsing 4 to monitor dew point in compressed air system. Using a rs232 to ethernet comverter to pull data into database. Works great.
Improvements to get 5 stars
1. Make the sensor replaceable by adding a plug
2. Replace power plug with a 5 pin m12 so you do not have to drill a hole in the housing to get serial communication out of unit.
I am trying to read through RS232. What is the command? How many Bytes supposed to return?
Asked by: KM OH
Hi KM OH,
Thank you for the inquiry. Please forward this question to TEMP@OMEGA.COM.
Best regards
OMEGA
Answered by: NorwalkAE_RS
Date published: 2024-02-22
What is the connector for 4 tp 20 mA ? M12?
Asked by: D LeRoy
Hi D LeRoy,
Thank you for the inquiry. This unit has lead wire with stripped end connection. We offer field mountable M12 connectors which can be used.
Best regards
OMEGA
Answered by: NorwalkAE_RS
Date published: 2024-02-04
Any examples of coding using RS232 protocol?
I want to collect data using LabVIEW. However, the sensor is updated every 5 seconds, and displaying an NO response error for the other 4 seconds. I want to see examples written in LabVIEW or other languages.
Asked by: KM OH
Thank you for the inquiry. Unfortunately we do not have any code examples for the RS232 protocol.
Best regards
OMEGA