Thermocouple input devices are instruments or components specifically designed to interface with thermocouple sensors and measure temperature using the thermoelectric effect. These devices enable the conversion of the voltage generated by the thermocouple junctions into temperature readings.
Examples of thermocouple input devices include:
- Thermocouple Amplifiers
- DAQ Systems
- Temperature Transmitters
- Programmable Logic Controllers (PLCs)
- Temperature Controllers
- Handheld Thermometers
Choosing a specific device or instrument depends on the application requirements, accuracy needs, measurement range, interface options, and other factors relevant to the temperature measurement system.
Erroneous Readings
There are a wide range of factors that can give a thermocouple measuring device an erroneous reading, and some common factors include:
- Temperature Gradient
- Cold Junction Compensation (CJC)
- Inaccurate Calibration
- Thermocouple Degradation
- Electrical Noise
- Connection Issues
- Thermocouple Wire Impurities
Short testing thermocouple input measuring devices can help determine if they are functioning correctly or contributing to measurement errors. This test involves creating a short circuit condition by connecting the positive and negative leads of the thermocouple together.
When short testing a thermocouple measuring device, the expected result is for the device to display a temperature reading of zero or close to zero. This test helps identify any potential issues with the device, such as faulty connections, sensor damage, or circuitry problems. If the device does not respond correctly during the short test, it suggests a problem with the measuring device itself or its calibration.
Short Testing in Action
If concerns about accuracy or reliability arise, short testing the thermocouple input device can help determine if it is functioning correctly or if it is contributing to measurement errors. Short testing involves connecting a piece of copper wire across the input terminals of the device to create a short circuit condition.
For example, if a thermocouple is connected to a plug-in card designed for thermocouple inputs and erratic readings are noticed, the short test can help identify potential issues. By removing the thermocouple and connecting a piece of copper wire across the input terminals (and connecting another piece from the negative terminal to the low-level ground), the device should ideally read room temperature and remain stable. If the readings are stable and show the ambient temperature when shorted, it indicates that the issue may not be with the card, PC, or software, but rather with the wiring or a broken thermocouple.
If the temperature input readings are still erratic during the short test, it may indicate a problem with the board itself, and it may need to be returned for repair. Testing the board with factory-supplied software or trying it on another PC can help narrow down the possibilities.
If the board works fine during the short test, then the thermocouple wiring must be checked. Moving the wiring around and using shielding techniques can help reduce noise and interference. In cases where electrical isolation is required, ungrounded thermocouples or signal conditioners can be used to provide isolation, noise filtering, signal amplification, and safety.
Short testing is a simple first step that can help narrow down the possibilities and identify potential issues with the thermocouple measuring device.