-30 to 1300°C, 16:1 Performance Infrared Thermometer
-30 to 1300°C, 16:1 FOV, Infrared Thermometer
Remises de volume disponibles
Calendrier de remise sur volume
Quantité | Prix unitaire |
---|---|
1-4 | C$254.10 |
5-9 | C$241.40 |
10-24 | C$236.31 |
25-49 | C$226.15 |
50-99 | C$221.07 |
100+ | C$215.99 |
Ajouté à votre panier
Configuration terminée.
Veuillez ajouter au panier pour la conserver ou quitter la configuration
Veuillez ajouter au panier pour la conserver ou quitter la configuration
Échec de l'enregistrement des configurations, cliquez sur le bouton Configurer Recommencer
- -50 to 1300°C (-58 to 2372°F) temperature range
- 16:1 Field of View
- Adjustable Emissivity: 0.10 to 1.00
- 500 ms response time
- “Hold” function retains the last reading
- High/Low Temperature Alarm
- Min/Max/Avg/Diff readings
- Single laser sighting for accurate targeting
- Get started with the quick start guide!
OS-1300-161 Options de modèles
Voir tous les modèlesModifiez ces spécifications pour commander un modèle différente. Toutes les combinaisons ne sont pas valides. Les options compatibles avec les sélections précédentes seront en gras.
*Les options en surbrillance ne sont pas compatibles. Veuillez sélectionner une combinaison différente.
Mesure de la température minimale
Mesure de température maximale
Laser Output
Valeur de précision
Résolution
Input Voltage
Spectral Response
Response Time
Champ de vision
The OS1300-Series handheld infrared thermometer is a non-contact temperature sensing thermometer. It has a wide temperature range of -50 to 1300°C (-58 to 2372°F). This thermometer features a backlit display which is automatically illuminated when the trigger is pulled and allows working in poor or zero-lighting conditions. The single laser is also automatically active when the trigger is pulled. The laser sight can be turned on and off with the press of a button. The laser is located above the IR sensor to locate the center of the area being measured and this series has a field of view of 16:1. The thermometer constantly updates the temperature measurement as long as the trigger is held in and will “hold” on the last temperature seen when the trigger is released. The OS1300-Series has adjustable emissivity, min, max, and averaging functions, as well as high and low alarm capability. This safely measures surface temperatures of hot, hazardous, or hard-to-reach objects without contact.
Accuracy Chart | ||
---|---|---|
Device Accuracy = ±3°C or ±3% of rdg, -30°C to 0°C (-22 to 32°F) | ||
Target Temperature | Upper Limit Reading | Lower Limit Reading |
5°C | 7°C | 3°C |
100°C | 102°C | 98°C |
150°C | 153°C | 147°C |
400°C | 408°C | 392°C |
1200°C | 1224°C | 1176°C |


RoHS 2015/863 - Voir le certificat
Approuvé CE