- Integral Temperature/Humidity Sensors
- Plus 2 External Inputs for Temperature (Pt100/Pt1000 RTD/Thermistor) or Temperature/Humidity Probes
- Up to 4 Million Readings (Internal Memory Size)
- Up to 4 Hz Sampling Rate
OMYL-RH25 Options de modèles
Voir tous les modèlesModifiez ces spécifications pour commander un modèle différente. Toutes les combinaisons ne sont pas valides. Les options compatibles avec les sélections précédentes seront en gras.
*Les options en surbrillance ne sont pas compatibles. Veuillez sélectionner une combinaison différente.
Measurement Type
Number of Measurements
Input Type
Maximum Samples
Power
Data Interface
Max Sampling Rate
Additional Feature
The OMYL-RH25 is a temperature/humidity data logger that also accepts two external temperature or temperature/humidity inputs. This data logger is available with standard memory of 100,000 readings (OMYL-RH25) or expanded memory of 4 million readings (OMYL-RH25-4M). The following inputs can be configured:
• Internal temperature and humidity sensors
• External temperature (Pt100/Pt1000 RTD, Type U thermistor)
• External temperature/humidity (for combined external temperature/humidity probes)
The storage medium is non-volatile flash memory so no data is lost if the battery becomes discharged. Data logger setup as well as data retrieval is accomplished using the included standard OMYL-SOFT software. Downloaded data is stored in the OMYL-SOFT software as a text file for viewing in third party software packages or Microsoft Excel. The OMYL-SOFT standard software does not include any graphing functionality. Graphing and analyzing downloaded data files requires the OMYL-SOFT-PLUS version of the software (sold separately).
SPECIFICATIONS
Internal Temperature
Temperature Range: -30 to 70°C (-22 to 158°F) with standard lithium battery; -40 to 90°C (-40 to 194°F) with optional high temperature lithium battery
Temperature Resolution: 0.01°C (0.02°F)
Accuracy: See chart on PDF
External Temperature Input
Sensor: Pt100 or Pt1000 RTD, U thermistor (2000Ω @ 25°C)
Range: -70 to 250°C (-94 to 482°F)
Resolution: 0.01°C (0.02°F)
Accuracy: See chart on PDF
Internal/External Humidity
Humidity Range: 0 to 100% RH
Humidity Resolution: 0.01% RH
Accuracy: See chart on PDF
Reading Rate: 1 second to 24 hours; 2 Hz, 4 Hz in fast mode
Start Modes: Immediate start, delayed start
Memory Capacity
OMYL-RH25: 100,000 readings
OMYL-RH25-4M: 4 million readings
Memory Wrap Around: Yes (software configurable)
Battery: 3V lithium battery [standard temperature range battery (included), optional high temperature range battery]; user replaceable
Battery Life (Standard Temperature Range Battery): 4 years at 1 minute reading rate, 230 days at 10 second reading rate, 25 days at 1 second reading rate
LED Functionality: Integrated status LED indicates data logging status, low battery or alarm condition
Alarms: Software programmable high/low thresholds
Computer Interface: USB (interface cable included)
Software: XP SP3/Vista/7 and 8 (32- and 64-bit)
Operating Environment: -30 to 70°C (-22 to 158°F) with standard lithium battery; -40 to 90°C (-40 to 194°F) with optional high temperature lithium battery
Dimensions: 30 H x 50 mm dia (1.18 x 1.97')
Weight: 70 g (2.5 oz)
Enclosure: Robust, shock-proof POM, IP65
• Internal temperature and humidity sensors
• External temperature (Pt100/Pt1000 RTD, Type U thermistor)
• External temperature/humidity (for combined external temperature/humidity probes)
The storage medium is non-volatile flash memory so no data is lost if the battery becomes discharged. Data logger setup as well as data retrieval is accomplished using the included standard OMYL-SOFT software. Downloaded data is stored in the OMYL-SOFT software as a text file for viewing in third party software packages or Microsoft Excel. The OMYL-SOFT standard software does not include any graphing functionality. Graphing and analyzing downloaded data files requires the OMYL-SOFT-PLUS version of the software (sold separately).
SPECIFICATIONS
Internal Temperature
Temperature Range: -30 to 70°C (-22 to 158°F) with standard lithium battery; -40 to 90°C (-40 to 194°F) with optional high temperature lithium battery
Temperature Resolution: 0.01°C (0.02°F)
Accuracy: See chart on PDF
External Temperature Input
Sensor: Pt100 or Pt1000 RTD, U thermistor (2000Ω @ 25°C)
Range: -70 to 250°C (-94 to 482°F)
Resolution: 0.01°C (0.02°F)
Accuracy: See chart on PDF
Internal/External Humidity
Humidity Range: 0 to 100% RH
Humidity Resolution: 0.01% RH
Accuracy: See chart on PDF
Reading Rate: 1 second to 24 hours; 2 Hz, 4 Hz in fast mode
Start Modes: Immediate start, delayed start
Memory Capacity
OMYL-RH25: 100,000 readings
OMYL-RH25-4M: 4 million readings
Memory Wrap Around: Yes (software configurable)
Battery: 3V lithium battery [standard temperature range battery (included), optional high temperature range battery]; user replaceable
Battery Life (Standard Temperature Range Battery): 4 years at 1 minute reading rate, 230 days at 10 second reading rate, 25 days at 1 second reading rate
LED Functionality: Integrated status LED indicates data logging status, low battery or alarm condition
Alarms: Software programmable high/low thresholds
Computer Interface: USB (interface cable included)
Software: XP SP3/Vista/7 and 8 (32- and 64-bit)
Operating Environment: -30 to 70°C (-22 to 158°F) with standard lithium battery; -40 to 90°C (-40 to 194°F) with optional high temperature lithium battery
Dimensions: 30 H x 50 mm dia (1.18 x 1.97')
Weight: 70 g (2.5 oz)
Enclosure: Robust, shock-proof POM, IP65