PST rejoint DwyerOmega. En savoir plus>
4 Channel K Type Thermocouple Meter NIST Traceable
4-Channel Type-K Thermometer with Backlit Display
Remises de volume disponibles
Calendrier de remise sur volume
Quantité | Prix unitaire |
---|---|
1-4 | C$253.12 |
5-9 | C$240.46 |
10-24 | C$235.40 |
25-49 | C$225.28 |
50-99 | C$220.21 |
100+ | C$215.15 |
Ajouté à votre panier
Configuration terminée.
Veuillez ajouter au panier pour la conserver ou quitter la configuration
Veuillez ajouter au panier pour la conserver ou quitter la configuration
Échec de l'enregistrement des configurations, cliquez sur le bouton Configurer Recommencer
- 4 Inputs with Differential Temperature Measurement
- 3 1 /2 Digital Display with Backlighting
- Measures K Thermocouples in °F or °C
- Read HOLD, MAX Function
HH501DK Options de modèles
Voir tous les modèlesModifiez ces spécifications pour commander un modèle différente. Toutes les combinaisons ne sont pas valides. Les options compatibles avec les sélections précédentes seront en gras.
*Les options en surbrillance ne sont pas compatibles. Veuillez sélectionner une combinaison différente.
Category
Input Type
Minimum Temperature Measurement
Maximum Temperature Measurement
Input Connection Type
Display Type
Accuracy Value
Resolution
Supply Power
Number of Channels
Features
The OMEGA™ HH501DK is a rugged, easy to use thermometer with 4 standard miniature connector inputs. It accepts K type thermocouple probes and covers ranges from -50 to 1300°C (-58 to 1999°F). It also provides differential temperature measurement readings of T1-T2, T1-T3, T1-T4, as well as individual readings of the 4 inputs.
Specifications
Temperature Range: -50 to 1300°C (-58 to 1999°F)
Input: Type K thermocouple
Accuracy:
±(0.3% rdg + 1°C) -50 to 1000°C
±(0.5% rdg + 1°C) 1000 to 1300°C
±(0.3% rdg + 2°F) -58 to 1999°F
Display: 3 1/2 digit liquid crystal display (LCD) with maximum reading of 1999
Polarity: Automatic, positive implied, negative polarity indication.
Overrange: (OL) or (-OL) is displayed
Zero: Automatic Low Battery Indication: visual indicator is displayed when battery voltage drops below operating level.
Measurement Rate: 2.5 times Operating Environment: 0 to 50°C @ <70% RH
Storage Temperature: -20 to 60°C, 0 to 80% RH with battery removed from meter
Accuracy: Stated accuracy @ 23°C ±5°C <75% relative humidity. (accuracy specification does not include type K thermocouple probe accuracy)
Dimensions: 195 x 92 x 53 mm (7.7 x 3.6 x 2.1')
Weight: Approx. 250 g (9 oz) including battery
Specifications
Temperature Range: -50 to 1300°C (-58 to 1999°F)
Input: Type K thermocouple
Accuracy:
±(0.3% rdg + 1°C) -50 to 1000°C
±(0.5% rdg + 1°C) 1000 to 1300°C
±(0.3% rdg + 2°F) -58 to 1999°F
Display: 3 1/2 digit liquid crystal display (LCD) with maximum reading of 1999
Polarity: Automatic, positive implied, negative polarity indication.
Overrange: (OL) or (-OL) is displayed
Zero: Automatic Low Battery Indication: visual indicator is displayed when battery voltage drops below operating level.
Measurement Rate: 2.5 times Operating Environment: 0 to 50°C @ <70% RH
Storage Temperature: -20 to 60°C, 0 to 80% RH with battery removed from meter
Accuracy: Stated accuracy @ 23°C ±5°C <75% relative humidity. (accuracy specification does not include type K thermocouple probe accuracy)
Dimensions: 195 x 92 x 53 mm (7.7 x 3.6 x 2.1')
Weight: Approx. 250 g (9 oz) including battery
Rated 5 out of
5
by
Jim Hightower from
Design problem
I used a shorting bar to check the four inputs to see if they were the same. They indicated the ambient temperature but, when the supplied thermocouples were plugged in, a different ambient temperature was indicated. I assume that this is due to the tolerance of the thermocouple wire. A bigger difficulty was the spacing of the TC receptacles. When all four TCs are plugged in, it is very difficult to remove one of the plugs. It was necessary for me to pry two of them out with a small screwdriver before the other two could be removed. This is a design problem that should never been approved for manufacture.
Date published: 2020-11-23
Does it come with a immersion thermocouple probe?
The probes that come with the HH501DK are not recommended for immersion and a metal sheathed probe would need to be purchased separately.
Date published: 2020-09-03
We just bought 2 of Omega HH501 Dk thermometers. When temping products it keeps going into OL while we are temping the product. Why does this happen?
The OL would be an open input so one of the Thermocouples is opening up intermittently or at specific temperatures. Make sure the probes are operating within their specified range and try others to filter out possible defective ones.
Date published: 2024-10-04
Does it have data output such as RS-232 or USB?
This unit does not provide a communications output.
Specifications are below:
Temperature Range: -50 to 1300°C (-58 to 1999°F)
Input: Type K thermocouple
Accuracy:
±(0.3% rdg + 1°C) -50 to 1000°C
±(0.5% rdg + 1°C) 1000 to 1300°C
±(0.3% rdg + 2°F) -58 to 1999°F
Display: 3 1/2 digit liquid crystal display (LCD) with maximum reading of 1999
Polarity: Automatic, positive implied, negative polarity indication.
Overrange: (OL) or (-OL) is displayed
Zero: Automatic
Low Battery Indication: visual indicator is displayed when battery voltage drops below operating level.
Measurement Rate: 2.5 times/sec
Operating Environment: 0 to 50°C @ <70% RH
Storage Temperature: -20 to 60°C, 0 to 80% RH with battery removed from meter
Accuracy: Stated accuracy @ 23°C ±5°C <75% relative humidity. (accuracy specification does not include type K thermocouple probe accuracy)
Dimensions: 195 x 92 x 53 mm (7.7 x 3.6 x 2.1")
Weight: Approx. 250 g (9 oz) including battery
Thank you for your question.
Date published: 2019-07-03
What is the length of the provided k thermocouples?
Thank you for your inquiry. The Type K Thermocouple 36" Long.
Date published: 2020-07-25
Does it come with type K thermocouples? Or, I have to buy thermocouples additionally. Is calibration needed before use, or it is already calibrated and tested before shipping? Thanks.
Hi, thanks for connecting with Omega. This unit comes complete with a NIST calibration certificate with no points, user’s manual, 4 beaded K Thermocouples, alkaline battery and a rubber boot. Please contact our Tech Support if you have any further questions. Thank you.
Date published: 2019-05-31