High Accuracy oil filled Wet/Dry and Wet/Wet Differential Pressure Transducers
0 to 1 psi, Differential, 4 to 20 mA, 1/4" NPT Male, Cable, (-49 to 240 °F)
Remises de volume disponibles
Calendrier de remise sur volume
Quantité | Prix unitaire |
---|---|
1-9 | C$1,497.83 |
10-24 | C$1,422.94 |
25-49 | C$1,392.98 |
50-99 | C$1,333.07 |
100+ | C$1,303.11 |
Ajouté à votre panier
Configuration terminée.
Veuillez ajouter au panier pour la conserver ou quitter la configuration
Veuillez ajouter au panier pour la conserver ou quitter la configuration
Échec de l'enregistrement des configurations, cliquez sur le bouton Configurer Recommencer
- FAST DELIVERY!
- 5-Point NIST Traceable Calibration
- Precision Micro-Machined Silicon Core
- High Stability, Low Drift
- Welded Stainless Steel Construction
- 316L SS Wetted Parts on Wet Side– Clean Dry Gases on Dry Side
- Now Available with M12 Connector
Order Model PX459
(See Part Number Builder at bottom of page) - Premium Temperature Performance
- Broad Compensated Temperature Range
- Durable, 1 Million Cycle Life
- Ruggedized with Secondary Containment
- Customized Specifications Available
PX409-001DDUI Options de modèles
Voir tous les modèlesModifiez ces spécifications pour commander un modèle différente. Toutes les combinaisons ne sont pas valides. Les options compatibles avec les sélections précédentes seront en gras.
* Les options en surbrillance ne sont pas compatibles. Veuillez sélectionner une combinaison différente ou cliquez sur le bouton CONFIGURER pour créer la vôtre.
Pressure Range
Accuracy Value(compensated range)
Output Signal
Electrical Output
Pressure Port Type
Connection/Termination
Operating Temperature Range
Additional Features
Pressure Type
Accuracy Class
Pressure Port Size
Get all the great performance characteristics of the Micro-Machined Silicon Transducers in these wet/wet differential pressure models. These rugged stainless steel transducers are ideal for tough industrial automotive, or aerospace applications. They can be used in test benches, fitler monitoring, hydraulic, flow, or water/wastewater applications, as well as factory or pneumatic air, pitot tubes, air speed and other industrial or aerospace applications. The solid state silicon core and avaliable all stainless steel wetted parts will provide long, reliable service life with excellent long term stability.


Rated 5 out of
5
by
Anonymous from
Just the Best
Since these USB-enabled transducers are performing perfectly in a unit operations lab at the University of Kansas, they are now installed in a new undergrad fluid mechanics lab experiment at the University of Delaware
Date published: 2020-04-30
RoHS 2015/863 - Voir le certificat
Approuvé CE
UKCA

A previous answer stated "A static pressure 500 psi above the rated differential pressure range... can be applied simultaneously to both ports." Does this apply to all models? (i.e. the 0-250dpsi model is rated to 750psi line pressure instead of 500psi?)
That would be correct
Date published: 2021-04-25
Should the transducer body be in a horizontal or vertical orientation when the ports are in a horizontal orientation?
The position of the body (electronics housing) should not matter. The orientation of the pressure ports (especially on the low differential range units) can effect the reading. It always best to scale the instrumentation that you using to reading the signal from your transducer with it oriented in the position you are going to be using it in.
Date published: 2021-09-10
Item# PX409-015DWUUSBH The data sheet states that the Output Signal is Digital and the Electrical Output is USB. Is there software available to monitor the transducers output if this connects thru USB?
Yes there is. Software can be downloaded free for Omega's website.
Date published: 2020-06-15
I'm attempting to get a steady state flow through one of our apparatuses for flow assurance research for glaucoma but there is a problem with air being in the sytem during the flow testing. What is the best way to eliminate air from the equation?
Tanner, thank you for your question, I do not have a specific operating procedure to eliminate air from the equation, since I am not familiar with your system design. Typically you will suction our Air or flush out air before you apply a positive pressure. Thank you.
Date published: 2019-11-16
Currently I have the black and red cables attached to a COM and V supply respectively. I was wondering of the remaining green and white cables, which one is used as the analog input wire?
If you are working with a 4 to 20mA output sensor only the Red and Black wire are used. In a 2-wire loop powered device the power supply, load and sensor are wired in series. If you are not familiar with this type of hook-up please contact Omega at 1-800-826-6342 and ask for the Pressure Group.
Date published: 2021-04-28
How can I find the response time of this transducer? (We need a fast response one)
Thank you for your inquire
These units have a under 1 mill second response time.
Be understand that these and all the transducers we carry are not dynamic products.
If you need more information? Please contact us directly.
Date published: 2022-07-15
At one side of the transducer, we are going to have an aqueous solution with fine solid particles. Is there any consideration for this condition? Are "oil filled" transducers proper for this fluid? If no, please introduce a porper type.
It depends on what you define as "fine" particles. The fitting has a small port that leads to a chamber where the diaphragm is located. The particles could possibly clog the port. This would not damage the unit but it would cause it to read incorrectly. If at all possible you should filter the solution to make it as clean as you can. The ideal solution would be a sensor with a flush diaphragm (no cavity). We do not offer a differential sensor with this feature
Date published: 2022-07-15
If 500 psi is the maximum line/static pressure that can be applied to both sides of the sensor simultaneously, how can these PX409s go up to a measurement range of 1000 psid?
On the 1000 PSID range units. A static pressure 500 psi above the rated differential pressure range (1500 PSI) can be applied simultaneously to both ports.
Date published: 2021-02-04